Delfield Scientific
For customer service, call (800) 733-8829, (800) 773-8821, Fax (989) 773-3210, www.delfield.com
SCIENTIFIC
Freezer Operation
Environmentally friendly R404A refrigerant is used. The
system has the capability of maintaining between -4°F (-20˚C)
and 0°F (-17˚C) in heavy use operations. System is controlled
using Delfield’s electronic temperature control, which provides
improved pull down times, reducing compressor cycling
and longer compressor life with lower energy consumption.
Control system automatically defrosts the evaporator coil
every 4 hours to assure evaporator coil is free of ice and
operating at optimum efficiency. Evaporator condensate is
eliminated using an energy efficient hot gas system.
With the doors closed, the temperature of the cabinet should
reach 0°F (-18°C) on freezers in about one hour. Continuous
opening and closing of the doors will hamper the unit’s ability
to maintain optimum temperature.
The electronic temperature control constantly monitors box
temperature as well as evaporator coil temperature to maintain
consistent product temperatures. As an added energy-saving
feature, the electronic controller will switch the evaporator
fan motor on and off with the compressor and condenser fan
motor.
Service Alert
During normal operation the evaporator fan may cycle and/
or pulse independently of the compressor. Contact Technical
Support at 1-800-733-8829 if you are unsure of the proper
function.
IMPORTANT NOTE REGARDING FREEZERS:
Whenever the freezer is plugged in, and the control has
completed initializing, the electronic temperature control will
cycle the compressor, evaporator fan motor, and condenser
fan motor to maintain box temperature at the control setting.
Freezer Automatic Defrost
The control also monitors compressor total running time
and will enter a defrost cycle after total compressor running
time is greater than 4-hours since the last defrost cycle OR
if evaporator coil temperature drops below -34°F (indicating
excessive frost on the coil).
Freezer Manual Defrost
When the control enters the defrost mode, whether manual
or automatic, it switches off the evaporator fan motor,
compressor and condenser fan motor, and switches on the
defrost heater to warm the evaporator coil and melt all frost
accumulated during the previous refrigeration cycle. The
control will continue the defrost cycle for a MINIMUM of 8
minutes and a MAXIMUM of 30 minutes depending on the
amount of frost accumulated on the evaporator coil.
After the defrost cycle is complete, the control returns to a
normal refrigeration cycle, however the evaporator fan motor
will not switch on for 2 minutes AFTER the compressor and
condenser fan motor have begun operating.
Electronic Temperature Control Location & Adjustment
The control is located in the control box at the rear of the unit.
It is factory set at mid-range to maintain about -3°F (-18°C)
box temperature.
Danfoss AK-CC210 Control
If at any time the wrong selection is made
stop pressing buttons and wait approximately 30
seconds and the display will time out. Restart
from the first step.
To Adjust Set Point (Desired Temp)
1. Press and release the center button one time and the
display will show the cabinet set point.
2. Use the top and bottom buttons to adjust the temp
higher or lower to the desired temperature. (note: if the
desired temperature is between 38 and 42 degrees set
the temperature set point at 40 as the temp will fluctuate
above and below the set point) .
3. Once the desired set point is reached on the display,
press and release the center button one time to accept
that temperature.
To Change From Celsius to Fahrenheit:
1. Press and hold the top button of the control until the
display shows r01 then release the button.
2. Press the bottom button approximately 4 times until the
display reads r05.
3. Press and release the center button on time and the
display will show ˚F or ˚C
4. To change use the top or bottom button to scroll and
select the temp format required.
5. Once the chosen format is visible in the display press and
release the center button one time.
6. The display will return and say r05.
7. Do not press any buttons at this point, The display will
time out in approximately 30 seconds and display the
chosen format.
The set point regulates air temperature. The display
is a measure of the bottle temperature. Initially there
could be up to 3°F (1.6°C) difference between these
two temperatures. It may take up to two hours for the
bottle temperature (display) to reach the set point.